| Measuring range |
0.01 µm - 4 mm |
| Measurement principle |
Laser Diffraction & Dynamic Image Analysis |
| Precision* |
Spherical glass beads D50 = 642 microns, precision as CV = 0.7% //Spherical glass beads D50 = 57 microns, precision as CV = 1.0% //Spherical latex beads D50 = 0.4 microns, precision as CV = 0.6% |
| Laser class |
Red 780 nm, blue 405 nm //Class 1 laser product per CFR 1040.10 & IEC60825-1 |
| Laser power |
Red laser 0.35 to 2 mW nominal //Blue laser 4-8 mW nominal |
| Detection system |
Two fixed photo-electric detectors with logarithmically spaced segments are placed at correct angles for optimal scattered light detection from 0 to 165 degrees using 151 detector segments. |
| Data |
Volume, number and area distributions as well as percentile and other summary data |
| Data format |
Stored in ODBC format in encrypted Microsoft Access Databases to ensure compatibility with external statistical software applications. |
| Data integrity |
Data integrity may be ensured using FDA 21 CFR Part 11 compliant security features including password protection, electronic signatures and assignable permissions |
| Type of analysis |
dry and wet analysis |
| Measuring time |
~ 10 to 30 seconds |
| Power requirements |
AC input: 90 - 264 VAC, 47 - 63 Hz, single phase |
| Power consumption |
25 W nominal, 50 W max., depending on options installed |
| Environmental conditions |
Temperature: 5° to 40° Celsius (50° to 95° Fahrenheit) //Humidity: 90% RH, non-condensing maximum //Storage Temperature: -10° to 50° Celsius (14° to 122° Fahrenheit) (dry only) //Pollution: Degree 2 |
| Standards |
Laser Diffraction (ISO 13320)//Dynamic Image Analysis (ISO 13322-2)//Representation of results of particle size analysis (ISO 9276-6) |
| Image analysis |
5.2 megapixel (2560 x 2048), 60 fps at max resolution |
| Wet operation |
Volume: 200 ml nominal //Flow rate: 0 to 65 ml/sec with water //Inlet pressure: 50 psig (345 kPa) maximum |
| Dry operation |
100 psi (689 kPa) maximum pressure //5 CFM (8.5 m3/h) at 50 psi (345 kPa) minimum flow rate //Free of dry contaminants, moisture and oil |
| Vacuum |
Vacuum must meet or exceed 50 CFM (85 m3/h) |
| Physical specifications |
Case Material: Impact resistant plastic //Exterior surfaces are finished with corrosion resistant paint or plating //Chemical Compatibility: Class I |
| Dimensions (W x H x D) |
~ 820 x 460 x 500 mm (32.3 x 18.1 x 19.7 in) |
| Weight (Measuring unit) |
FlowSync: 19.5kg (43lbs)//TurboSync: 13.6kg (30lbs)//Sync: 23.6kg (50.8lbs) |