Mercury porosimeter BELPORE Series Technical Data
| Function | LP: Degassing, Hg filling, low-pressure porosimetry//MP & HP: High-pressure porosimetry |
|---|---|
| Pressure range | LP: Vacuum up to 450 KPa//MP: 0.1 - 228 MPa//HP: 0.1 - 414 MPa |
| Resolution LP | 0.001 kPa in measuring range: vacuum to 0.1 kPA / 0.01 kPa in measuring range 0.1 - 450 kPA |
| Resolution MP | 0.001 MPa in measuring range 0.1 - 100 MPa / 0.01 MPa in measuring range 100 - 228 MPa |
| Resolution HP | 0.001 MPa in measuring range 0.1 - 100 MPa / 0.01 MPa in measuring range 100 - 414 MPa |
| Pressure detection accuracy | better than 0.1% F.S. |
| Pore size diameter | LP: 180 - 3.25 µm (UMP: 1,000 - 3.8 µm)//MP: 15 - 0.0065 µm//HP: 15 - 0.0036 µm |
| Particle size diameter | LP: 330 - 15 µm (UMP: 3,000 - 15 µm)//MP: 40 - 0.015 µm//HP: 40 - 0.01 µm |
| Max. detectable volume | LP & MP: 0.5 cm³ - 2 cm³//HP: 0.5 cm³ |
| Volume detection accuracy | better than 1% F.S. |
| Max. number of measuring points | 10.000 intrusion//10.000 extrusion |
| Weight | LP: 55 kg (121 lbs)//MP & HP: 68 kg (150 lbs) |
| Dimensions (W x D x H) | LP: 40 x 67 x 80 cm//MP & HP: 40 x 67 x 80 cm |